CNV 508N4-series 4对8线耦合去耦网络 IEC60255-26
符合 IEC 60255-26 标准的 4 对 8 线耦合 去耦网络
Application
标准
Benefits
Electrical Diagram
技术数据
(PDF-文档 6.2 MB)
Highlights:
Selection of all coupling modes as per IEC 60255-26
Remote controlled by the OCS 500N6 N6F
Coupling capacitor 0.5uF
Decoupling inductor >1.5mH per line
Nominal line voltage max. 250V ACDC
Nominal line current 4A ACDC or 16A ACDC
Application:
CouplingDecoupling networks for testing measuring relays and protection equipmentThe CNV 508N4 series are special couplingdecoupling networks being used to perform "Electrical disturbance tests for measuring relays and protection equipment" in accordance to IEC 60255-26. A capacitance of 0.5uF is specified to couple the pulses on to the lines.
Each line has a decoupling inductor of >1.5mH to protect the auxiliary equipment from being subjected to the test pulse. The disturbance signal used for this test is a damped oscillatory pulse with a frequency of 1MHz, generated by the EM TEST
OCS 500N6 or OCS 500N6F generators.
标准:
IEC 60255-22-1
IEC 60255-26
Benefits:
Fully-automatic coupling networks for IEC 60255-26Depending on the configuration of the DUT several different test set-up's are used to perform tests in differential mode (line-to-line of the same pair) or mon mode (each pair to ground or one pair to all other pairs with these being grounded). The CNV 508N4 is automatically set by the OCS 500N6 selecting the desired coupling mode and lines to be coupled. The coupler offers capability to test up to 4 pairs8 lines.
The test of measuring relays and protection equipment as outlined in IEC 60255-26 includes different coupling modes to test the DUT under different conditions. Each pair of lines is tested individually in differential mode. Common mode test is applied to each pair to ground and to each pair with the other pairs connected to ground.
The EM TEST CNV 508N4 series are a dedicated CDNs for these applications, fully automatic, remote controlled by the OCS 500N6 for the various coupling modes. This offers best user convenience and reliability that all couplings required are tested sequently. The selected coupling mode and lines actually being tested are indicated by LEDs.
Electrical Diagram:
The plete solution
The CNV 508N4 series includes the required 1.5mH decoupling inductors to decouple the auxiliary equipment as well as the coupling part via a 0.5uF capacitor for each line and the coupling relays to allow all possible settings for differential mode and mon mode testing as per
IEC 60255-26. The coupling mode and the coupling of the lines are selected and controlled via the OCS 500N6.